ELISA Troubleshooting Guide

Having difficulty with your ELISA?

There may be an easy solution to the problem. Review the information shown below to find some common causes for errors while running an ELISA test and the possible fixes. 

 

Download PDF

PROBLEM

POSSIBLE CAUSE

SOLUTION/REMEDY

Weak or No Signal

Reagents added/prepared incorrectly

Check protocol, ensure that the reagents were added in the proper order and prepared to correct dilution.

Reagents not equilibrated to room temperature

Per specific product protocol ensure appropriate reagents equilibrate to room temperature prior to starting the assay.

Incorrect storage of reagents/Expired Reagents

Double check the expiration dates and storage conditions in product protocol and/or component labels.

Very low incubation temperature or agitation

Control your room (lab) temperature to bring it within the recommended range (Refer to the Package Insert). Set your plate shaker to 600rpm.

Wells scratched with pipette or washing tips

Use caution when dispensing and aspirating into and out of wells. Automated plate washers may need to be calibrated so tips don't touch the bottom of the wells.

Plate washing too forceful

Check and ensure correct pressure in automatic plate washer. Pipette wash buffer gently if manually washing the plate.

Plate read at incorrect wavelength

Double check product protocol to ensure plate reader is set to the correct wavelength.

Incompatible sample type

Detection may be reduced or absent in untested sample types. Include a sample that the assay is known to detect as a positive control.

High Background Across the Plate
Standard Curve is Saturated
Too Much Signal

Insufficient washing

Wash wells as per the product protocol. As a general rule, increasing the duration of the soak step may help (Add 30 seconds to the soak time). In addition, at the end of each washing step, invert the plate on absorbent tissue and tap forcefully if necessary to remove any residual fluid.

Contaminated wash buffer

Prepare a fresh solution of wash buffer.

Wrong conjugate dilution used

Dilute the conjugate at the recommended dilution.

The assay was incubated for too long in one or all steps

Strictly follow the assay incubation time in all steps according to the procedure.

Substrate incubation not protected from light

Substrate incubations are commonly performed in the dark. Review the product protocol and follow recommendation of the manufacturer.

The substrate is contaminated (not fresh)

Check the color of the substrate - it should be colorless.

High incubation temperature

Control your room (lab) temperature to bring it within the recommended range (Refer to the Package Insert).

Dirty plate/well bottom

Clean the bottom of the plate.

PROBLEM

POSSIBLE CAUSE

SOLUTION/REMEDY

Positive Signal (high background) in Negative Controls or Standard Wells

Contaminated Pipette tips

Use clean pipette tips and change tips for every sample and/or standard.

Cross contamination has occurred between wells

Carefully wash wells. The wash step is very critical.

Incorrect standard dilutions prepared

Review protocol and double check calculations to ensure proper dilution of standards/controls.

Poor Standard Curve in ELISA

Incorrect standard curve dilutions prepared

Confirm dilutions are made correctly and avoid cross contamination of standards by utilizing new pipette tips with each standard preparation.

Standard improperly reconstituted

Briefly spin vial prior to opening. Inspect vial for undissolved material after reconstitution.

Possible degradation of standard material

Ensure storage and handling of the standard was properly followed as recommended in the product protocol.

Curve doesn't fit data

Try plotting standard data using a different scale.
e.g. log-log, 4 Parameter logistic, 5 parameter logistic curve fits.

High CV across the plate (high variation in samples / standards)

Pipetting problem

When using:

Single Channel Pipette: Check tips for bubbles between replicates.

Multichannel Pipette: Calibrate the pipette. Check tips for bubbles before dispensing.

Bubbles in wells

Ensure no bubbles are present prior to reading the plate.

Edge Effects

Ensure the plate and all reagents are at room temperature.

Insufficient washing

Wash wells as per the product protocol. As a general rule, increasing the duration of the soak step may help (Add 30 seconds to the soak time). In addition, at the end of each washing step, invert the plate on absorbent tissue and tap forcefully if necessary to remove any residual fluid.

Non-Uniform Washing

All the wells should be uniformly washed. Check your washing system and/or method.

Inconsistent sample storage

Ensure sample storage conditions are maintained per guidelines outlined in the assay protocol.

Non-homogeneous reagents, samples, or standards

Ensure all reagents are mixed thoroughly. Vortex samples/standards to ensure homogeneity before pipetting.

Sample Readings are out of the assay dynamic range

Sample contains undetectable analysis level, i.e. below assay limit of detection (LOD)

If samples fall below the LOD, contact us: info@ibl-america.com.

Sample contains high analyte level: i.e. above assay highest standard point

Samples should be diluted and re-assayed.

 

Pipetting Technique Tips for ELISA

  1.  Use only pipettes within the range suggested by the manufacturer.
  2.  Make sure disposable pipette tip is fit tightly onto the pipette.
  3.  Check that there are no air bubbles while pipetting reagents/samples/standards/controls.
  4.  Reduce splashing of liquid by pipetting into the sides of the microplate wells.
  5.  Use a new pipette tip between pipetting of standards, controls, and samples.

 